  • Rangedown: up to 100: 1 (5: 1 range turndown)
  • Measurement range: up to 1,000 inH2O (2.48 bar) differential
  • Pipe diameter: 50 to 200 mm (2 to 8 in.)
  • Output: differential pressure, specific variable
  • Communication protocol: WirelessHART®
  • Material in contact with the process: 316 stainless steel
  • Basic diagnostics
  • Certifications / Approvals: NACE®, hazardous areas; for the entire list of certifications, see the complete specifications
  • Refresh rate: user selectable, from 1 s to 60 min
  • Power Module Life: Up to 10 year life, field replaceable (order separately)
  • Wireless range: indoor antenna (225m)